Surface temperature

Surface temperature

T56

P/N: 765500 [View large]
T56 is a Pt-100 surface temperature sensor that connects to SwemaTemp 20.
Measuring uncertainty to some degree, can occure due to heat transfer between surface, sensor and shaft.
T56 connects to SwemaTemp 20.

Surface -40...300°C
pt-100
4-wire
Surface Ø 6 x 150 mm
T90=45sec

Uncertainty
(95% coverage probability)
-30...70°C: ±0,3°C
70...300°C: ±(0,3°C + 0,3% read value)
-40...-30°C: ±0,4°C
Uncertainty is valid after correction is made according to calibration cert. At room temp. max: ±0,4°C
Uncertainty due to heat transfer shall be added (method uncertainty)