Temperature sensor for surfaces.
Swa 56 fits any Swema 3000 model.
Measuring uncertainty to some degree, can occure due to heat transfer between surface, sensor and shaft.
SWA 56 connects to Swema 3000.
Surface -40...300°C
pt-100
4-wire
Surface Ø 6 x 150 mm
T90=45sec
Uncertainty
(95% coverage probability)
0-50°C: ±0,1°C
50...300°C: ±0,2% read value
-40...0°C: ±(0,1°C + 0,2% read value)
Uncertainty due to heat transfer shall be added (method uncertainty)